Ultra-high sensitive acetylene detection using quartz-enhanced photoacoustic spectroscopy with a fiber amplified diode laser and a 30.72 kHz quartz tuning fork

Date
2017
Journal Title
Journal ISSN
Volume Title
Publisher
AIP Publishing LLC
Abstract

An ultra-high sensitive acetylene (C2H2) Quartz-enhanced photoacoustic spectroscopy (QEPAS) sensor based on a high power laser and a quartz tuning fork with a resonance frequency f0 of 30.72 kHz was demonstrated. An erbium-doped fiber amplifier (EDFA) amplified distributed feedback diode laser with a center wavelength of 1.53 μm was used as the exciting source. A 33.2 ppb minimum detection limit (MDL) at 6534.37 cm−1 was achieved, and the calculated normalized noise equivalent absorption coefficient was 3.54 × 10−8 cm−1 W/√Hz when the laser output power was 1500 mW. The ppb-level detection sensitivity of C2H2 validated the reported QEPAS method.

Description
Advisor
Degree
Type
Journal article
Keywords
Citation

Ma, Yufei, He, Ying, Zhang, Ligong, et al.. "Ultra-high sensitive acetylene detection using quartz-enhanced photoacoustic spectroscopy with a fiber amplified diode laser and a 30.72 kHz quartz tuning fork." Applied Physics Letters, 110, no. 3 (2017) AIP Publishing LLC: http://dx.doi.org/10.1063/1.4974483.

Has part(s)
Forms part of
Rights
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Link to license
Citable link to this page