A NDIR Mid-Infrared Methane Sensor with a Compact Pentahedron Gas-Cell
dc.citation.articleNumber | 5461 | en_US |
dc.citation.issueNumber | 19 | en_US |
dc.citation.journalTitle | Sensors | en_US |
dc.citation.volumeNumber | 20 | en_US |
dc.contributor.author | Ye, Weilin | en_US |
dc.contributor.author | Tu, Zihan | en_US |
dc.contributor.author | Xiao, Xupeng | en_US |
dc.contributor.author | Simeone, Alessandro | en_US |
dc.contributor.author | Yan, Jingwen | en_US |
dc.contributor.author | Wu, Tao | en_US |
dc.contributor.author | Wu, Fupei | en_US |
dc.contributor.author | Zheng, Chuantao | en_US |
dc.contributor.author | Tittel, Frank K. | en_US |
dc.date.accessioned | 2020-10-21T18:16:22Z | en_US |
dc.date.available | 2020-10-21T18:16:22Z | en_US |
dc.date.issued | 2020 | en_US |
dc.description.abstract | In order to improve the performance of the large divergence angle mid-infrared source in gas sensing, this paper aims at developing a methane (CH4) sensor with non-dispersive infrared (NDIR) technology using a compact pentahedron gas-cell. A paraboloid concentrator, two biconvex lenses and five planar mirrors were used to set up the pentahedron structure. The gas cell is endowed with a 170 mm optical path length with a volume of 19.8 mL. The mathematical model of the cross-section and the three-dimension spiral structure of the pentahedron gas-cell were established. The gas-cell was integrated with a mid-infrared light source and a detector as the optical part of the sensor. Concerning the electrical part, a STM32F429 was employed as a microcontroller to generate the driving signal for the IR source, and the signal from the detector was sampled by an analog-to-digital converter. A static volumetric method was employed for the experimental setup, and 20 different concentration CH4 samples were prepared to study the sensor’s evaluation, which revealed a 1σ detection limit of 2.96 parts-per-million (ppm) with a 43 s averaging time. | en_US |
dc.identifier.citation | Ye, Weilin, Tu, Zihan, Xiao, Xupeng, et al.. "A NDIR Mid-Infrared Methane Sensor with a Compact Pentahedron Gas-Cell." <i>Sensors,</i> 20, no. 19 (2020) MDPI: https://doi.org/10.3390/s20195461. | en_US |
dc.identifier.doi | https://doi.org/10.3390/s20195461 | en_US |
dc.identifier.uri | https://hdl.handle.net/1911/109441 | en_US |
dc.language.iso | eng | en_US |
dc.publisher | MDPI | en_US |
dc.rights | This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited | en_US |
dc.rights.uri | https://creativecommons.org/licenses/by/4.0/ | en_US |
dc.subject.keyword | infrared spectroscopy | en_US |
dc.subject.keyword | mid-infrared sensor | en_US |
dc.subject.keyword | NDIR | en_US |
dc.subject.keyword | pentahedron gas-cell | en_US |
dc.title | A NDIR Mid-Infrared Methane Sensor with a Compact Pentahedron Gas-Cell | en_US |
dc.type | Journal article | en_US |
dc.type.dcmi | Text | en_US |
dc.type.publication | publisher version | en_US |