Feasibility of On-line Proton Beam Range Verification with Positron Emission Tomography Imaging

dc.contributor.advisorClark, Jr., John W.en_US
dc.contributor.committeeMemberKelly, Kevin F.en_US
dc.contributor.committeeMemberJacot, Jeffret G.en_US
dc.contributor.committeeMemberShao, Yipingen_US
dc.creatorLou, Kaien_US
dc.date.accessioned2016-01-25T16:45:19Zen_US
dc.date.available2016-01-25T16:45:19Zen_US
dc.date.created2016-05en_US
dc.date.issued2015-12-18en_US
dc.date.submittedMay 2016en_US
dc.date.updated2016-01-25T16:45:19Zen_US
dc.description.abstractPositron emission tomography (PET) imaging has been adopted clinically to verify proton beam range (BR) in proton therapy. Conventional approaches use off-line verification that verifies BR after delivery of full dose of a treatment fraction. This verification can be used to check the accuracy of current therapy retrospectively, but its effect suffers from potential target deviation between treatment fractions. In contrast, on-line verification can verify BR using a low dose proton beam and a short data acquisition time prior to administration of a treatment fraction. It could provide adequate information to verify or even revise the beam delivery for improving therapy targeting within a single treatment fraction. Nevertheless, a practical on-line verification has many unknowns and technical challenges. This thesis studies the feasibility of on-line BR verification with PET imaging from three aspects. First, it investigates the impact of count statistics on the accuracy and precision of BR verification. In general, on-line verification is hampered by low count statistics that degrade PET images and subsequently the accuracy and precision of BR verification. We used Monte Carlo simulation to understand the quantitative relationship between count statistics and the accuracy and precision of BR verification under various proton irradiation and PET imaging conditions. A mathematical model was also developed to study the impact of Poisson noise associated with PET images on BR verification. Second, it evaluates the effectiveness of existing algorithms and develops new approaches for improving the accuracy and precision of on-line BR verification. Important factors were identified, effective and efficient approaches were developed. By optimizing and applying these factors and approaches, it is feasible to substantially improve the accuracy and precision of on-line BR verification with the same proton dose and negligible extra processing time. Lastly, it investigates the feasibility of achieving on-line BR verification for proton brain therapy with Monte Carlo simulation, and proposes a new PET system configuration for on-line verification. It demonstrates that by using proper data correction and advanced algorithms, it is feasible to achieve 1-mm accuracy and precision of on-line BR verification with low proton dose and short data acquisition under different proton irradiation conditions.en_US
dc.format.mimetypeapplication/pdfen_US
dc.identifier.citationLou, Kai. "Feasibility of On-line Proton Beam Range Verification with Positron Emission Tomography Imaging." (2015) Diss., Rice University. <a href="https://hdl.handle.net/1911/88105">https://hdl.handle.net/1911/88105</a>.en_US
dc.identifier.urihttps://hdl.handle.net/1911/88105en_US
dc.language.isoengen_US
dc.rightsCopyright is held by the author, unless otherwise indicated. Permission to reuse, publish, or reproduce the work beyond the bounds of fair use or other exemptions to copyright law must be obtained from the copyright holder.en_US
dc.subjectProton Therapyen_US
dc.subjectPETen_US
dc.subjectOn-line Measurementen_US
dc.titleFeasibility of On-line Proton Beam Range Verification with Positron Emission Tomography Imagingen_US
dc.typeThesisen_US
dc.type.materialTexten_US
thesis.degree.departmentApplied Physicsen_US
thesis.degree.disciplineNatural Sciencesen_US
thesis.degree.grantorRice Universityen_US
thesis.degree.levelDoctoralen_US
thesis.degree.majorApplied Physics/Electrical Engen_US
thesis.degree.nameDoctor of Philosophyen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
LOU-DOCUMENT-2016.pdf
Size:
6.94 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 2 of 2
No Thumbnail Available
Name:
PROQUEST_LICENSE.txt
Size:
5.84 KB
Format:
Plain Text
Description:
No Thumbnail Available
Name:
LICENSE.txt
Size:
2.6 KB
Format:
Plain Text
Description: