Analog Computer Simulation of Maximum Expiratory Flow Limitation

Date
1975-01-20
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract

This paper extends previous modeling work in our laboratory on the simulation of the panting maneuver executed in a body plethysmograph [Golden, et al. (1973)] to the simulation of a wider range of pulmonary function tests and, in particular, the maximum expiratory flow-volume (MEFV) curve. the simulation is implemented on an analog computer for ease of parameter manipulation; the effects of changes in several model parameters (such as large and small airway resistence, and airway and lung compliance on the MEFV curve are discussed. The model structure presented here is much simpler than the ventilatory sytem models of Fry (1968) and Pardaens, et al. (1972) and yet yields the same information in a diagnostic sense.

Description
Tech Report
Advisor
Degree
Type
Report
Keywords
phethysmograph, pulmonary
Citation

M. F. Olender, J. J. Clark and P. Stevens, "Analog Computer Simulation of Maximum Expiratory Flow Limitation," Rice University ECE Technical Report, no. TR7508, 1975.