Roberts, Jabus B.2009-06-042009-06-041988Nguyen, Chau T.. "Calibration of a leadglass lead-scintillator photon detector at Fermilab experiment E-704 (Lead-scintillator)." (1988) Master’s Thesis, Rice University. <a href="https://hdl.handle.net/1911/13383">https://hdl.handle.net/1911/13383</a>.https://hdl.handle.net/1911/13383We calibrated a leadglass Pb-scintillator photon detector with a positron beam of 30 GeV. The detector consisted of a segmented leadglass counter in front and a Pb-scintillator sandwich counter in the back. The response of the sandwich counter exhibited an exponential dependence on the position of the incident particle. The calibration constants of the leadglass segments were obtained by an iterative method. The results of our analysis showed an energy resolution of $\sigma ({\rm E})\over {\rm E}$ = 24%/$\surd$E. A monitoring system using LED and $\sp{241}$Am sources regularly checked the stability of all modules. A typical photomultiplier's gain factor deviated no more than 6% from the mean value at any time during a period of three months.87 p.application/pdfengCopyright is held by the author, unless otherwise indicated. Permission to reuse, publish, or reproduce the work beyond the bounds of fair use or other exemptions to copyright law must be obtained from the copyright holder.Particle physicsElementary particlesHigh energy physicsCalibration of a leadglass lead-scintillator photon detector at Fermilab experiment E-704 (Lead-scintillator)ThesisThesis Phys. 1989 Nguyen